Confotec CARS 系统
多种同步分析功能:
高速 CARS信号成像
共聚焦拉曼成像
共聚焦荧光成像,包含双光子荧光(或多光子)
共聚焦激光反射成像
高对比度激光透射成像
三种类型CARS测试:
F-CARS
E-CARS
P-CARS
五独立通道同步获取4种2D及3D图像
F-CARS
E-CARS及拉曼
激光反射成像
激光透射成像
激光激发成像
激发及探测偏振控制
计算机控制全自动测试模式切换
三种扫描模式
XY阵镜扫描
XY自动位移台扫描
XY阵镜扫描+自动位移台扫描实现高分辨率全景成像
Spatial XYZ resolution of CARS signals (Objective lens: 60х, NA=1.2, watter immersion): | < 0.7 μm |
Spatial resolution of Raman signals (excitation wavelength 633 nm; objective lens: 60х, NA=1.2, watter immersion): | XY: < 300 nm Z: < 700 nm |
Spectral detectable range: | CARS signals: 985 – 5000 cm-1 |
Spectral resolution: | CARS signals: 7-8 cm-1 Raman signals: 0.25 cm-1(grating 75 l/mm Echelle); 0.6 cm-1 (grating 1800 g/mm) |
Scanning range (fast scanning mode, 60x lens) | XY: 225 х 225 μm Z: 80 μm |
Control and automation: | Fully motorized |
Type: | inverted |
Model: | Nikon Ti-U |
Stage: | motorized |
- travel range: | 114 х 75 mm |
- accuracy (for 1 mm traveling): | 0.06 μm |
- XY repeatability: | ± 1 μm |
- minimal step: | 0.02 μm |
Micro- objective lens: | 60 х NA-1.2 water immersion 20 x NA-0.45 |
Z-scanner: | |
- type: | piezo scanner |
- lens movement range: | 80 μm |
- minimal step: | 50 nm |
- Reapetability: |